{"ID":2835605,"CreatedAt":"2026-06-01T04:54:23.091178241Z","UpdatedAt":"2026-06-01T04:54:23.091178241Z","DeletedAt":null,"paper_url":"https://arxiv.org/abs/2511.23355","arxiv_id":"2511.23355","title":"A Hierarchical Computer Vision Pipeline for Physiological Data Extraction from Bedside Monitors","abstract":"In many low-resource healthcare settings, bedside monitors remain standalone legacy devices without network connectivity, creating a persistent interoperability gap that prevents seamless integration of physiological data into electronic health record (EHR) systems. To address this challenge without requiring costly hardware replacement, we present a computer vision-based pipeline for the automated capture and digitisation of vital sign data directly from bedside monitor screens. Our method employs a hierarchical detection framework combining YOLOv11 for accurate monitor and region of interest (ROI) localisation with PaddleOCR for robust text extraction. To enhance reliability across variable camera angles and lighting conditions, a geometric rectification module standardizes the screen perspective before character recognition. We evaluated the system on a dataset of 6,498 images collected from open-source corpora and real-world intensive care units in Vietnam. The model achieved a mean Average Precision (mAP@50-95) of 99.5% for monitor detection and 91.5% for vital sign ROI localisation. The end-to-end extraction accuracy exceeded 98.9% for core physiological parameters, including heart rate, oxygen saturation SpO2, and arterial blood pressure. These results demonstrate that a lightweight, camera-based approach can reliably transform unstructured information from screen captures into structured digital data, providing a practical and scalable pathway to improve information accessibility and clinical documentation in low-resource settings.","short_abstract":"In many low-resource healthcare settings, bedside monitors remain standalone legacy devices without network connectivity, creating a persistent interoperability gap that prevents seamless integration of physiological data into electronic health record (EHR) systems. To address this challenge without requiring costly ha...","url_abs":"https://arxiv.org/abs/2511.23355","url_pdf":"https://arxiv.org/pdf/2511.23355v1","authors":"[\"Vinh Chau\",\"Khoa Le Dinh Van\",\"Hon Huynh Ngoc\",\"Binh Nguyen Thien\",\"Hao Nguyen Thien\",\"Vy Nguyen Quang\",\"Phuc Vo Hong\",\"Yen Lam Minh\",\"Kieu Pham Tieu\",\"Trinh Nguyen Thi Diem\",\"Louise Thwaites\",\"Hai Ho Bich\"]","published":"2025-11-28T17:02:27Z","proceeding":"cs.CV","tasks":"[\"cs.CV\"]","methods":"[]","has_code":false}